. 2015 Nov 5;105:401–411. doi: 10.1007/s00392-015-0934-0

Table 3.

Procedural features

DIOR (n = 20) IN.Pact Falcon (n = 25) p value
Predilatation with standard balloon 20 (100) 24 (96) 0.37
Predilatation balloon diameter, mm 3.1 ± 0.4 2.9 ± 0.4 0.07
Predilatation balloon length, mm 17.4 ± 5.4 20.7 ± 7.9 0.26
Predilatation pressure, ATM 14.8 ± 4.5 14.2 ± 5.0 0.71
>1 DEB used per lesion 2 (10) 6 (24) 0.27
DEB diameter, mm 3.2 ± 0.4 3.1 ± 0.3 0.34
 2.50 2 (10) 1 (4)
 2.75 0 3 (12)
 3.00 9 (45) 13 (52)
 3.50 8 (40) 8 (32)
 4.00 1 (5) 0
DEB length, mm 27.3 ± 9.4 32.4 ± 12.0 0.10
DEB inflation pressure, ATM 10.8 ± 3.0 11.3 ± 3.0 0.54
DEB inflation time, s 58.0 ± 5.3 52.0 ± 13.2 0.18
Postdilatation with standard balloon 3 (15) 1 (4) 0.31
Maximum balloon diameter to index stent diameter ratio 1.09 ± 0.08 1.07 ± 0.09 0.34
Additional stenting 1 (5) 2 (8) 0.69
Angiographic success 20 (100) 25 (100)
Device success 18 (90) 23 (92) 0.69
Procedural success 20 (100) 25 (100)