. 2015 Nov 5;105:401–411. doi: 10.1007/s00392-015-0934-0

Table 6.

Percentage changes between postprocedure and follow-up

DIOR (n = 20) IN.pact falcon (n = 25) Crude p value Adjusted p value*
Quantitative coronary angiography 18 (90) 23 (92)
 Minimal lumen diameter change, % –10.4 [–42.3 to 4.7] 4.8 [−7.7 to 16.8] 0.034 0.26
 Diameter stenosis change, % 28.6 [−7.1 to 187] 2.6 [−49.7 to 67.6] 0.032 0.32
Optical coherence tomography 17 (85) 21 (84)
 Minimal lumen area change, % −30.2 [−49.5 to 2.6] −13.4 [−21.8 to 19.4] 0.048 0.097
 Maximal neointimal area change, % 33.8 [6.2 to 72.0] −8.9 [−21.0 to 33.0] 0.002 0.007
 Maximal neointimal area stenosis change, % 35.8 [8.7 to 59.1] 14.6 [−21.7 to 36.6] 0.014 0.009
 Lumen volume change, % −14.6 [−34.0 to 1.7] 2.89 [−14.0 to 18.6] 0.011 0.026
 Stent volume change, % −0.7 [−3.0 to 2.7] −1.6 [−6.9 to 5.9] 0.67 0.77
 Neointimal volume change, % 27.2 [1.1 to 58.6] −15.8 [−36.7 to 28.3] 0.006 0.028
Fractional flow reserve 17 (85) 22 (88)
 FFR stent gradient change, % 69.0 [0.0 to 238] −40.8 [−58.9 to 18.8] 0.003 0.46

* Adjusted for differences in hypercholesterolemia, postdilatation and postprocedure OCT minimal lumen area