Skip to main content
. 2016 Apr 12;2016:3849316. doi: 10.1155/2016/3849316

Table 2.

Mean duration of core vitrectomy operating time for 27-gauge and 25-gauge surgeries utilizing either a standard single-cut or a TDC vitrectome.

Parameter 25-gauge vitrectomy 27-gauge vitrectomy
Standard vitrectome (n = 20) TDC vitrectome (n = 20) Standard vitrectome (n = 20) TDC vitrectome (n = 20)
Mean core vitrectomy duration (SD), in seconds 148.19 (25.14) 73.80 (18.61) 242.71 (25.52) 159.71 (29.47)
p value: significance of comparison between core vitrectomy durations performed using a standard and a two-dimensional cutting (TDC) vitrectome (t-test) p < 0.001 p < 0.001

Mean duration of core vitrectomy operating time for 27-gauge and 25-gauge surgeries utilizing either a standard single-cut or a TDC vitrectome is shown above, demonstrating significantly shorter case duration for TDC vitrectomy surgery groups (n = 80, 20 eyes assigned, resp., to each surgery group).