. 2016 Apr 15;16(4):542. doi: 10.3390/s16040542

Table 7.

Electrocardiogram signal quality—saltwater immersion with movement condition.

Parameter Ag/AgCl PLS p-Value Ag/AgCl PLP p-Value
Temporal measures of HRV meanNN(ms) 1454 ± 223 938 ± 124 0.0020 1277 ± 355 851 ± 120 7.11e-04
SDNN (ms) 1102 ± 315 497 ± 190 0.0054 872 ± 412 335 ± 243 7.92e-04
RMSSD (ms) 1481 ± 309 677 ± 265 0.0023 1197 ± 558 448 ± 293 4.53e-04
NN50 132.2 ± 41.2 117.6 ± 22.9 0.5072 171.8 ± 116.5 102.1 ± 58.5 0.0785
Spectral measures of HRV LF (ms2) 296431 ± 35372 45736 ± 29809 0.1529 174843 ± 241251 26346 ± 30832 0.0459
HF (ms2) 531535 ± 37302 146587 ± 10063 0.0564 439754 ± 499125 57801 ± 58326 0.0151
Total (ms2) 2984351 ± 1396 1088667 ± 4357 0.0199 2459872 ± 20369 748602 ± 261051 0.0085
LF/HF 0.4661 ± 0.2187 0.3257 ± 0.0298 0.1928 0.4170 ± 0.1328 0.4440 ± 0.1370 0.6284
Corr-Coef r (unitless) 0.21 ± 0.14 0.51 ± 0.17 0.0000 0.21 ± 0.14 0.63 ± 0.13 0.0000