Table 5. Measured dynamic range values in the scintillator decay scans, and values of K and D obtained from the model.
| Scintillator | Synchrotron | Bunch mode | Energy (keV) | Maximum | Minimum | Dynamic range | K | D |
|---|---|---|---|---|---|---|---|---|
| CRY-019 | DLS | 686 | 55 | 42.0 | 3.11 | 13.5 | 1.621 | 4.18 |
| LYSO:Ce | DLS | 686 | 55 | 56.9 | 2.88 | 19.8 | 2.349 | 3.36 |
| YAG:Ce | DLS | 686 | 55 | 10.6 | 2.66 | 3.98 | 0.086 | 2.90 |
| LuAG:Ce | ESRF | 4BM | ID19 U17-6c beam | 7.12 | 4.76 | 1.50 | 2.251 | 3.72 |
| LYSO:Ce | ESRF | 4BM | ID19 U17-6c beam | 36.3 | 1.84 | 19.7 | 33.45 | 2.63 |
| LYSO:Ce | ESRF | 16BM | ID19 U17-6c beam | 578 | 34.4 | 16.8 | 556.9 | 21.4 |