FIG. 4.

Platelets were repeatedly exposed to various shear stresses and exposure times, with PAS measured for the resulting stress accumulation. Significance was achieved for stress accumulations above 1611 dyne-s/cm2, or 74 passages at 350 dyne/cm2 for 50 ms (*p < 0.001).