FIG. 5.

FIG. 5

PABST saturation levels versus total off-resonance (in ppm) for a constant FOV with six different arrays. Plots display the fraction of fat pixels with Mz < 0.1 following PABST pulses optimized for the different coils. Little gain is seen for channel counts greater than 12. At extremely high off-resonance (8 and 10 ppm), the PABST method is able to effectively saturate over three times more fat than a single-channel transmitter. [Color figure can be viewed in the online issue, which is available at wileyonlinelibrary.com.]