Skip to main content
. 2016 May 6;6:25465. doi: 10.1038/srep25465

Table 1. The difference between glucometer (No. 1–11) and laboratory method (No. 12) in retro-orbital blood glucose levels.

  Glucose range in No. 12 (mg/dL)
0–99 100–199 200–299 300–
(mg/dL) (%) (mg/dL) (%) (mg/dL) (%) (mg/dL) (%)
Glucose value (No. 1–12) (mg/dL) 20.6 ±  3.4 34.7 ±  6.6 43.5 ±  8.8 30.5 ±  5.7 53.8 ±  11.1 20.1 ±  4.0 65.1 ±  13.6 16.5 ±  3.3
Glucose value (No. 2–12) (mg/dL) 30.1 ±  2.6 47.0 ±  5.3 42.9 ±  8.0 30.2 ±  5.2 60.5 ±  14.0 22.9 ±  4.9 101.6 ±  21.9 25.2 ±  5.3
Glucose value (No. 3–12) (mg/dL) − 9.2 ±  3.3 − 10.3 ±  4.5 10.8 ±  8.3 7.4 ±  5.4 8.5 ±  9.1 3.0 ±  3.3 37.6 ±  19.1 8.5 ±  4.5
Glucose value (No. 4–12) (mg/dL) − 7.7 ±  3.5 8.1 ±  5.0 10.8 ±  8.3 7.5 ±  5.4 8.3 ±  7.6 2.9 ±  2.8 35.1 ±  20.1 7.7 ±  4.8
Glucose value (No. 5–12) (mg/dL) 6.1 ±  3.2 11.9 ±  5.0 30.7 ±  8.7 21.5 ±  5.6 38.0 ±  9.7 14.2 ±  3.5 26.2 ±  10.9 6.8 ±  2.8
Glucose value (No. 6–12) (mg/dL) − 8.5 ±  3.9 − 9.5 ±  5.3 − 11.3 ±  7.1 − 7.3 ±  4.5 − 26.6 ±  9.3 − 10.0 ±  3.3 − 52.4 ±  14.2 − 13.6 ±  3.7
Glucose value (No. 7–12) (mg/dL) − 0.5 ±  3.1 2.3 ±  4.6 16.3 ±  8.9 11.3 ±  5.7 28.3 ±  11.0 10.5 ±  4.1 53.1 ±  19.7 12.8 ±  4.8
Glucose value (No. 8–12) (mg/dL) 13.0 ±  3.4 22.3 ±  5.5 29.0 ±  8.6 20.2 ±  5.6 37.7 ±  9.8 14.4 ±  3.6 48.0 ±  14.1 12.0 ±  3.4
Glucose value (No. 9–12) (mg/dL) 3.6 ±  3.2 8.7 ±  5.0 18.2 ±  8.8 13.1 ±  5.7 22.8 ±  10.2 8.7 ±  3.7 39.3 ±  13.7 9.7 ±  3.4
Glucose value (No. 10–12) (mg/dL) − 1.1 ±  3.2 8.7 ±  4.6 23.8 ±  9.5 16.2 ±  6.0 42.8 ±  10.8 16.3 ±  3.9 64.2 ±  18.5 16.0 ±  4.7
Glucose value (No. 11–12) (mg/dL) 27.2 ±  3.5 44.1 ±  6.6 72.3 ±  11.3 49.7 ±  7.4 109.6 ±  13.9 41.3 ±  4.8 130.0 ±  14.0 33.5 ±  3.7

values are mean ±  SE. Sample number at − 100 mg/dL, 100–200 mg/dL, 200–300 mg/dL and 300– mg/dL glucose range is 29, 33, 13 and 10 respectively.