Table 1. Manufacturer-specific acquisition details.

Acquisition parameter Philips-specific OVS Siemens-specific OVS Philips-specific ZOOM Siemens-specific 2DRF
TR (ms)* ~7000 ~7000 ~7000 ~7000
TE (ms) 72 89 50 81
Matrix size 100x100 100x100 64x48 64x38
Bandwidth (Hz) 893.1 1085 893.1 1100
SENSE (Philips) or GRAPPA (Siemens) acceleration factor 1.5 2.0 1.5 2.0
Partial Fourier? No 6/8 No 7/8
Number of diffusion-weighted directions 32 30 32 30
Number of b = 0 acquisitions 4 4 4 4

*Times approximate due to cardiac gating