. 2001 Feb 1;106(1):105–149. doi: 10.6028/jres.106.006

Table 17.

Expanded uncertainty (k = 2) in radiance temperature for the Cs or Na pressure-controlled heatpipe blackbody source at 800 °C. The uncertainty is given for two different target diameters of the radiation thermometer

Factor Radiation thermometer target diameter (mm)
1.0 10
BB Emissivity 0.02 0.02
Au/Pt thermocouple 0.01 0.01
Digital voltmeter & ice bath 0.00 0.00
Pressure stability 0.02 0.02
RT noise 0.02 0.02
BB radiance uniformity 0.06 0.57
Total 0.07 0.57