Table 18.

Expanded uncertainty (k = 2) in radiance temperature for an argon-filled ribbon filament lamp in the RTCL. The uncertainty, in °C, is given for different radiance temperatures (also in °C)

Source of Uncertainty Type Radiance temperature (°C)
800 1100 1500 1900 2300
Calibration of the reference radiance temperature lamp relative to the 1990 NIST Radiance Temperature Scale B 0.12 0.19 0.32 0.48 0.67
Test lamp temperature determination A 0.42 0.17 0.29 0.43 0.60
Lamp current measurement B 0.29 0.19 0.15 0.14 0.14
Mean effective wavelength measurement for the NIST PEP B 0.10 0.04 0.09 0.28 0.54
Test lamp alignment B 0.09 0.14 0.24 0.36 0.51
1990 NIST Radiance Temperature Scale relative to thermodynamic temperature scale B 0.15 0.24 0.40 0.61 0.85
Overall uncertainty of test lamp calibration with respect to SI units B 0.6 0.4 0.7 1.0 1.5