. 2001 Feb 1;106(1):105–149. doi: 10.6028/jres.106.006

Table 19.

Expanded uncertainty (k = 2) in radiance temperature for a typical radiation thermometer. The uncertainty, in °C, is given for different radiance temperatures (also in °C)

Source of Uncertainty Type Radiance temperature (°C)
800 1100 1500 1900 2300
Calibration of the variable temperature blackbody relative to the 1990 NIST Radiance Temperature Scale B 0.2 0.3 0.4 0.6 0.8
Mean effective wavelength measurement for the NIST PEP B 0.1 0.0 0.1 0.3 0.5
Blackbody uniformity B 0.2 0.2 0.2 0.2 0.2
Test thermometer temperature determination A 0.6 0.6 0.6 0.6 0.6
1990 NIST Radiance Temperature Scale relative to thermodynamic temperature scale B 0.1 0.2 0.4 0.6 0.9
Overall uncertainty of test radiometer calibration with respect to SI units B 0.7 0.8 0.9 1.1 1.5