Skip to main content
. 2001 Feb 1;106(1):105–149. doi: 10.6028/jres.106.006

Table 3.

Capsule standard platinum resistance thermometer ITS-90 calibrations. Vapor pressure is denoted by VP and the expanded uncertainty (k = 2) is denoted by U

ITS-90 Fixed Points e-H2 TP e-H2 VP e-H2 VP Ne TP O2 TP Ar TP Hg TP H2O TP Ga MP In FP Sn FP
ITS-90 assigned temperature (K) 13.8033 17.0 20.3 24.5561 54.3584 83.8058 234.3156 0.01 302.9146 429.7485 505.078
ITS-90 subranges U
(mK)
U
(mK)
U
(mK)
U
(mK)
U
(mK)
U
(mK)
U
(mK)
U
(mK)
U
(mK)
U
(mK)
U
(mK)
Max U
(mK)
13.8033 K to 273.16 K 0.22 0.21 0.22 0.26 0.18 0.08 0.20 0.02 0.62
24.5561 K to 273.16 K 0.22 0.26 0.18 0.08 0.20 0.02 0.39
54.3584 K to 273.16 K 0.18 0.08 0.20 0.02 0.29
83.8058 K to 273.16 K 0.08 0.20 0.02 0.39
234.3156 K to 302.9146 K 0.20 0.02 0.04 0.20
273.15 K to 302.9146 K 0.02 0.04 0.04
273.15 K to 429.7485 K 0.02 0.10 0.10
273.15 K to 505.078 K 0.02 0.10 0.24 0.24