Table 2.
The present and expected relative expanded uncertainties (coverage factor k = 2) for calibrations of spectral irradiance E(λ) at typical wavelengths
Wavelength (nm) | Present expanded uncertainty (%) | Expected expanded uncertainty (%) |
---|---|---|
250 | 1.8 | 1.0 |
350 | 1.1 | 0.6 |
655 | 0.9 | 0.5 |
900 | 1.1 | 0.5 |
1600 | 1.4 | 0.5 |
2400 | 4.4 | 1.1 |