Skip to main content
. 2001 Feb 1;106(1):151–186. doi: 10.6028/jres.106.007

Table 2.

The present and expected relative expanded uncertainties (coverage factor k = 2) for calibrations of spectral irradiance E(λ) at typical wavelengths

Wavelength (nm) Present expanded uncertainty (%) Expected expanded uncertainty (%)
250 1.8 1.0
350 1.1 0.6
655 0.9 0.5
900 1.1 0.5
1600 1.4 0.5
2400 4.4 1.1