Table 1.
Numerically computed and experimental results comparing the ERC-2L and the ERC-ML coils.
| ERC-2L | ERC-ML | ||
|---|---|---|---|
| Simulated | B1+ Transmit Efficiency (μT/W0.5) | 2.01 ± 1.35 | 2.04 ± 1.48 |
| Worst-case peak 1g SAR (W/kg) | 10.9 | 16.7 | |
| B1+ SAR Efficiency (μT/SAR0.5) | 0.61 | 0.50 | |
| SNR (a.u.) | 1.46 ± 0.92 | 1.51 ± 1.07 | |
| Experimental | B1+ Transmit Efficiency (μT/W0.5) | 1.53 ± 0.45 | 1.64 ± 1.00 |
| SNR (×105 Hz0.5/ml) | 5.33 ± 2.56 | 5.41 ± 2.98 |