. 2016 Apr 13;58(3):186–192. doi: 10.3164/jcbn.15-61

Table 1.

Precision studies for LC-MS/MS and two immunoassays performed using same two serum pools with different 25(OH)D concentrations

Within-run Between-run
Mean ± SD (nmol/L) CV (%) Mean ± SD (nmol/L) CV (%)
Centaur pool 1 132.0 ± 13.6 10.3 130.4 ± 14.3 11
pool 2 338.0 ± 31.1 9.2 340.2 ± 17.3 5.1
Elecsys pool 1 14.0 ± 0.2 1.4 14.3 ± 3.6 25.2
pool 2 33.6 ± 0.5 1.5 36.2 ± 7.1 19.6
LC-MS/MS 25(OH)D2 pool 1 18.7 ± 0.5 2.7 19.3 ± 1.1 5.7
pool 2 44.4 ± 1.0 2.3 46.5 ± 2.4 5.2
LC-MS/MS 25(OH)D3 pool 1 17.9 ± 0.4 2.2 18.3 ± 0.8 4.4
pool 2 38.9 ± 1.3 3.3 37.9 ± 1.7 4.5