Table 3.
Passing–Bablok regression and concordance correlation analysis for each immunoassay against LC-MS/MS total or 25(OH)D2
Passing–Bablok regression |
Concordance correlation |
|||||||||
---|---|---|---|---|---|---|---|---|---|---|
Intercept | 95%CI | Slope | 95%CI | Cusum test | CCC | 95%CI | r | Cb | ||
Before-Centaur | 16.9 | 10.8 to 22.1 | 0.48 | 0.34 to 0.65 | p>0.05 | 0.43 | 0.25 to 0.58 | 0.52 | 0.83 | |
Before-Elecsys | −5.3 | −19.4 to 3.6 | 1.12 | 0.87 to 1.47 | p>0.05 | 0.69 | 0.52 to 0.80 | 0.69 | 0.99 | |
After-Centaur | −13.8 | 4.4 to 19.8 | 0.53 | 0.42 to 0.73 | p>0.05 | 0.39 | 0.25 to 0.51 | 0.68 | 0.58 | |
After-Elecsys | 13.8 | −22.9 to 1.9 | 1 | 0.78 to 1.22 | p>0.05 | 0.61 | 0.45 to 0.73 | 0.75 | 0.81 | |
Adjusted 25(OH)D2 | ||||||||||
Centaur | −8.4 | −13.1 to −3.7 | 1.05 | 0.73 to 1.50 | p>0.05 | 0.34 | 0.17 to 0.49 | 0.54 | 0.63 | |
Elecsys | −19.1 | −30.8 to 10.9 | 1.54 | 1.07 to 2.50 | p>0.05 | 0.26 | 0.09 to 0.42 | 0.41 | 0.64 |
CCC, concordance correlation coefficient; CI, confidence interval; r, Pearson’s correlation coefficient; Cb, bias correction factor; Before-Centaur, Centaur values before treatment; Before-Elecsys, Elecsys values before treatment; After-Centaur, Centaur values after treatment; After-Elecsys, Elecsys values after treatment.