Skip to main content
. 2016 May;55(3):331–335.

Table 1.

Summary statistics for implantable microchip and rectal thermometry

Thermometry method Temperature (°F)
Mean difference between methods 95% Agreement limits (°F)
Range Average
Microchip 99.5–103.0 101.4 0.069 −1.82 to +1.96
Calibrated rectal 99.4–103.2 101.4 Not applicable
Microchip 99.2–103.2 101.4 −0.675 −2.19 to +0.84
Common rectal 99.0–102.6 100.7 Not applicable