Skip to main content
. 2016 Feb 19;7(3):911–942. doi: 10.1364/BOE.7.000911

Table 2. OCT angiography scan protocols.

Parameter Value
Pixels per B-scan 1536
B-scan repeats (B-scans per MB-scan) 5
Number of MB-scans (slow scanner positions) 180
Total number of B-scans 900
Number of A-scans per B-scan 300
Number of scanner fly-back A-scans (not acquired) 100
A-scan acquisition time 10 µs
B-scan acquisition time (time interval for OCTA data analysis) 4 ms
Volume acquisition time 3.6 s
Data size ~830 MB
 
Subject code N1 N2 N3 GA
 
Scanned retinal eccentricity 6° nasal, 4°inferior 8° nasal, 4°inferior 8° nasal, 4°inferior 3° nasal, 3°inferior
Scanned area 1.8 x 1.8 mm 1.8 x 1.3 mm 1.8 x 1.3 mm 1.6 x 1.6 mm
Horizontal step size 6 µm 6 µm 6 µm 5 µm
Vertical step size 10 µm 7 µm 7 µm 9 µm