Skip to main content
. 2016 May 13;27(3):90–97. doi: 10.7171/jbt.16-2703-001

TABLE 1.

Differences in Termini Counts among Different Micrometers per Pixel and Image Processing Iterations

Line number A
B
C
D
E
F
G
H
Lines/frame at scanning Target particle size (pixel2) Original image (counts) Median filter (counts) Optimized method (counts) Gaussian filter (sigma values) Percent of counts relative to highest resolution Relative data file size
1 8192 ≥320 641 558 470 100 100% 74×
2 4096 ≥80 935 584 522 100 111% 18.7×
3 2048 ≥20 1456 578 508 50 108% 4.6×
4 950 ≥4 1915 986 552 50 117%
5 512 ≥1 2307 485 765 25 163% 0.29×
6 256 ≥1 685 70 494 25 105% 0.07×
7 128 ≥1 353 11 248 25 53% 0.02×

Seven images, representing an identical tissue area of 775 × 775 µm, were acquired using a 20×/0.7 NA objective lens with varying sampling rates. Images were acquired stepwise with an ∼4-fold change in pixel area at each step. For every frame size, the lower limit of the particle size was defined according to the micrometer/pixel size of the image, and the cutoff was set at 2 µm where applicable. Line 4 represents the optimal lateral resolution, based on the structures of interest. Column A, Sampling rate shown as frame size; column B, particle size used for quantification; columns C–E, quantification results across several background processing iterations [column C, original image, no processing; column D, application of a median filter (radius = 2); column E, application of the proposed method]; column F, sigma radius values for the “generate blurred image” step; column G, percentage of counts in column E compared with the highest resolution image (line 1, column E); column H, difference in file size compared with the optimized setting. Note the increase in false counts with inappropriate processing when pixel size approaches target structure size.