Table 2.
Sensitivity of the different methods determined as the lowest PC concentration exhibiting a response above the threshold.
| Technology used | ELISA | MSD | Gyrolab | AlphaLISA | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Acid dissociation step | No | Yes | No | Yes | No | Yes | No | Yes | ||||||||
| Target depletion step | No | Yes | No | Yes | No | Yes | No | Yes | No | Yes | No | Yes | No | Yes | No | Yes |
| Final dilution applied before platform analysis | 1 in 6 | 1 in 45 | 1 in 6 | 1 in 45 | 1 in 6 | 1 in 9 | 1 in 15 | 1 in 45 | ||||||||
| Sensitivity ng/mL of reference ADA | 5 | 100 | 2 | 20 | 2 | 5 | 20 | 20 | ||||||||