Comparative table of different photo-detectors used in PET [13]

PMT APD SiPM
Gain 106 50–1000 106
Rise time (ns) ~1 ~5 ~1
QE @ 420 nm (%) ~25 ~70 ~25–75(PDE)
Temperature sensitivity (%/°C) <1 ~3 1–8
Bias (V) >1,000 300–1,000 30–80
Magnetic field sensitivity yes no no
Sensitive area cm2 mm2 mm2