Table 3.

Summary of parameters to consider in selection of an assay technology

FA TR-FRET EFC AlphaLisa/AlphaScreen SPA
Radiation No No No No Yes
Read-out Fluorescence Time gated emission Chemilunescence Fluorescence Scintillation
Scalability, max 1536 1536 1536 1536 384
Standard plate readers Yes Yes Yes Yes No
Compound interference High, low with red dyes Low Low Medium High, low with red beads and plates
Sensitivity Medium Medium High High High
Robustness, Z-factor 0.5–0.75 0.7–0.8 >0.7 >0.8 0.5–0.8