Table 7. Bending of implant neck and torque around stem.

HIGH100 data at the instance of peak force Fres. Mbn = bending moment at base of neck (Fig 1) without friction, only caused by Fres. Mts = torque around stem without friction. Mbs = bending moment in stem without friction. ΔMbn, ΔMts and ΔMbs = changes due to additionally acting friction moments at head. ISO Neck/Stem = ISO 7206-6/4 test forces.

Activity → Moments Cycling Sit Down Stand Up Knee Bend Walking Stance Stairs Up Stairs Down Jogging ISO Stem ISO Neck
Mbn [Nm] in Neck 24.3 54.4 59.8 52.8b 59.3 69.8 71.2 81.5a 107.1 (56.7) 131.7
ΔMbn [%] -0.7 1.2 0.5 1.5a 0.5 0.8 0.0b 0.6 0.3 --- ---
Mts[Nm] in Stem 15.5 22.9 19.3 15.6 24.8 11.7b 47.0a 32.4 57.4 16.0 (37.0)
ΔMts [%] 1.0 -1.8b 2.9 1.7 0.8 7.2a 0.9 -0.3 -0.4 --- ---
Mbs [Nm] in Stem 36.8 57.0 46.4 36.8b 65.9 55.4 113.1a 90.2 149.1 69.6 (161.5)
ΔMbs [%] 2.7 -3.3a 5.2 5.5a 0.6 -1.0 0.0 0.2 0.2 --- ---

a/b = highest/lowest absolute values during activities except cycling and jogging.