Table 1.
Dynamic range calculations for MST data.
B*
= 10 nM; signal/noise = 50 |
||||||||
---|---|---|---|---|---|---|---|---|
KD (nM) | ||||||||
0.3 | 0.4 | 0.5 | 0.6 | 0.7 | 0.8 | 0.9 | 1.0 | |
| ||||||||
successful trialsa | 8,301 | 9,566 | 9,898 | 9,975 | 9,998 | 10,000 | 10,000 | 10,000 |
| ||||||||
mean value | 0.3 | 0.4 | 0.5 | 0.6 | 0.7 | 0.8 | 0.9 | 1.0 |
| ||||||||
C.I.b | [0.2, 0.4] | [0.3, 0.5] | [0.4, 0.7] | [0.5, 0.8] | [0.5, 0.9] | [0.6, 1.0] | [0.7, 1.1] | [0.8, 1.2] |
| ||||||||
C.I. (ESP)c | [0.1, 0.6] | [0.2, 0.7] | [0.3, 0.8] | [0.3, 1.0] | [0.4, 1.1] | [0.5, 1.2] | [0.6, 1.3] | [0.7, 1.4] |
| ||||||||
B* = 2 nM; signal/noise = 10 | ||||||||
| ||||||||
successful trials | N.D. | N.D. | 7,206 | 8,155 | 8,742 | 9,051 | 9,368 | 9,621 |
| ||||||||
mean value | - | - | 0.6 | 0.7 | 0.8 | 0.9 | 1.0 | 1.1 |
| ||||||||
C.I.b | - | - | [0.2, 0.9] | [0.3, 1.1] | [0.4, 1.2] | [0.4, 1.4] | [0.5, 1.5] | [0.6, 1.6] |
| ||||||||
C.I. (ESP)c | - | - | [0.2, 1.8] | [0.2, 2.0] | [0.3, 2.2] | [0.3, 2.4] | [0.3, 2.6] | [0.4, 2.8] |
out of 10,000 total trials
Confidence interval calculated from the histogram of fitted values
Confidence interval calculated as the respective means of ESP limits for all successful trials