Table 3.

Dose differences for phantom study split in comparison for titanium and plastic applicator for organ at risk and target surrogate. A negative difference means the value was higher for TG-43 than for Acuros-based treatment plans

Absorbed dose EQD2 (α/β = 3 resp. 10 Gy)
TG-43
(Gy)
Acuros
(Gy)
relΔ
(%)
TG-43
(GyISO)
Acuros
(GyISO)
relΔ
(%)
OAR D0.1cm3 Titanium 5.32 5.22 –1.88 8.85 8.58 –3.06
Plastic 4.95 4.92 –0.61 7.87 7.79 –0.98
D2cm3 Titanium 4.43 4.34 –2.03 6.58 6.37 –3.22
Plastic 4.23 4.20 –0.71 6.12 6.05 –1.12
Target D50 Titanium 11.43 11.28 –1.24 20.40 20.02 –1.89
Plastic 11.25 11.24 –0.02 19.91 19.90 –0.04
D90 Titanium 7.54 7.44 –1.24 11.02 10.82 –1.76
Plastic 7.51 7.47 –0.51 10.96 10.88 –0.73
D98 Titanium 6.79 6.71 –1.11 9.49 9.35 –1.56
Plastic 6.77 6.77 –0.05 9.47 9.46 –0.07
V100 (%) (%) (%)
Titanium 96.37 95.35 –1.02
Plastic 96.14 95.75 –0.39

EQD2 – equivalent dose at 2 Gy, α/β – alpha/beta ratio, D0.1cm3, D2cm3 – minimum dose to the most exposed 0.1 cm3, 2 cm3, OAR – organs at risk, D50, D90, D98 – the minimum dose to 50%, 90%, 98% of the target, V100 – volume of the anatomic volume receiving 100%, of the prescribed dose