. 2016 May 5;18(5):e85. doi: 10.2196/jmir.5414

Table 2.

Systolic blood pressure measurements (mmHg) taken by the reference monitor and Checkme in the supine and upright positions.

Volunteers’ position Meana SD Range (min; max)a
Supine position



Reference monitor 136.6 21.8 84.7 (106.3; 191.0)

Checkme at heart level 138.4 25.2 94.5 (94.5; 189.0)

Checkme above heart 130.7b 27.7 101.0 (86.0; 187.0)
Upright position



Reference monitor 139.2 22.3 100.7 (102.3; 203.0)

Checkme at heart level 136.6c 25.9 87.7 (102.3; 190.0)

aThe average or range of 3 consecutive blood pressure measurements.

b P<.001 compared with Checkme at heart level.

c P=.01 compared with Checkme at heart level in the supine position.