Skip to main content
. Author manuscript; available in PMC: 2017 Jun 1.
Published in final edited form as: J Orthop Res. 2015 Dec 10;34(6):1019–1025. doi: 10.1002/jor.23109

Table 1.

Summary characteristics of the 18 test specimens from lab prepared (9 LP, n=9 donors) and postmortem retrieval TKRs (9 PM specimens from n=7 donors, 1 donor had bilateral implants, 2 specimens from one donor).

Donor Data Type Mean Standard Dev Min Max
Age (years) LP 76.7 8.7 64 90
PM 72.9 8.7 61 87
Weight (kg) LP 90.1 18.0 65 117
PM 78.3 17.7 61 103
Time in service (years) LP 0 0 0 0
PM 6.5 4.8 1 16
Supporting Bone Morphology & Mechanics
Supporting bone fraction (BV/TV) LP 0.122 0.054 0.072 0.236
PM 0.115 0.042 0.065 0.173
Supporting bone strain (με) LP −1320 545 −2020 −520
PM −2030 1570 −4820 −90
Interlock Morphology
Initial interdigitation depth, inID (mm) LP 2.3 0.7 1 3.2
PM 1.7 1 0.3 3.0
Current interdigitation depth, curID (mm) LP 2.3 0.7 1 3.2
PM 1.1 1.2 0 3.0
Gap thickness (μm) LP 14.0 4.1 7.1 21.2
PM 40.5 17.8 14.1 60.3
Interlock Mechanics
Trabeculae-cement micromotion (μm) LP 1.03 0.61 0.22 2.17
PM 5.67 4.71 0.1 13.1
Interdigitated trabecular bone strain (με) LP −410 330 −1060 −16
PM −1380 1530 −4300 15
Interdigitated trabecular bone strain (με) with distance from cement border > 0.25 mm LP −270 290 −830 52
PM −66 370 −500 550