Table 3.

Comparison of the performance of the biosensor modified either with PVA or BSA as a blocking agent

Blocking agent Sensitivity [Ω decade-1] Linear range [orders of magnitude] FET/ASF signal ratio a RSD [%]
BSA 790 ± 40 3 1.5 ± 0.2 8.4 ± 5.8
PVA 550 ± 20 7 3.5 ± 0.5 4.6 ± 3.7
a

– calculated by dividing the sensitivity of the device for FET to the sensitivity of the biosensor for ASF

All measurements were performed with three independent biosensor devices