. 2016 May 23;54(6):1660–1663. doi: 10.1128/JCM.00445-16

TABLE 1.

Sensitivity and Specificity of direct ESwab PCR and Lim broth PCR with the BD Max GBS assaya

Comparison ESwab PCR, %
Lim broth PCR, %
ESwab PCR vs Lim broth PCR, %
McNemar test
Estimate 95% CIb Estimate 95% CI Difference 95% CI
Sensitivityc
    Same-day ESwab vs Lim broth PCR 92.7 (101/109) 86.2–96.2 99.1 (108/109) 95.0–99.8 −6.4 −12.8 to −1.4 0.0156
    24-h ESwab vs Lim broth PCR 97.2 (106/109) 92.2–99.1 99.1 (108/109) 95.0–99.8 −1.8 −6.7 to 2.2 0.5000
Specificityd
    Same-day ESwab vs Lim broth PCR 96.7 (291/301) 94.0–98.2 95.0 (286/301) 91.9–97.0 1.7 −0.1 to 3.8 0.0625
    24-h ESwab vs Lim broth PCR 96.0 (289/301) 93.2–97.7 95.0 (286/301) 91.9–97.0 1.0 −0.8 to 3.1 0.375
a

Lim broth culture is the reference method.

b

95% CI, 95% confidence interval.

c

Values in parentheses represent the ratio of positive results to true positives.

d

Values in parentheses represent the ratio of negative results to true negatives.