Fig 3. FACSPresto performance with EQA material.

Fig 3

NHLS external quality assessment (EQA) scheme samples (n = 20) was analysed on the FACSPresto and SDI values (dark solid line) calculated for (a) absolute CD4 counts and (b) CD4% of lymphocytes. C and D indicate the corresponding performance of PLG/CD4 on the MPL system used as the predicate method in this study. The radar graphs depict the optimal target of zero (dashed line) and the acceptable range of -2 to 2 SDI (red circles with grey shaded area).