Table 2.
Characteristic | System B | System P |
---|---|---|
LS spectrometer model | Beckman LS7800 | Packard Tri-carb A2500TR |
Operating mode | Sum-coincidence | Sum-coincidence |
Photomultiplier tubes | Hamamatsu R331-05 | Hamamatsu R331-08 |
Operating temperature | Ambient | Ambient |
Coincidence resolving time | 22 ns | 18 ns |
Sum-coincident pulse amplification | Logarithmic | Linear |
Pulse resolving time | 5 μs to 33 μs (variable with pulse height) |
12μs (fixed) |
Spectral analog-to-digital converter (ADC) capacity | 1000 channels | 2048 channels |
Nominal conversion gain (energy per channel) | Variable (with logarithmic energy) |
≈1 keV |
Detection threshold (nominal) | ≤ 1 keV | ≤ 1 keV |
Live-time determination method (and uncertainty) | Gated oscillator (scaled) (± 0.1 %) |
Gated oscillator (scaled) (± 0.1 %) |
Quench indicating parameter (QIP) | Horrocks number (H) | Transformed Spectral Index of the External Standard (tSIE) (proprietary) |
External (-ray source for QIP determination | 137Cs | 133Ba |
(and location) | (side) | (bottom) |