. 1997 Jul-Aug;102(4):455–477. doi: 10.6028/jres.102.031

Table 2.

Characteristics of the NIST LS spectrometers employed in the calibration of 63Ni SRM 4226C

Characteristic System B System P
LS spectrometer model Beckman LS7800 Packard Tri-carb A2500TR
Operating mode Sum-coincidence Sum-coincidence
Photomultiplier tubes Hamamatsu R331-05 Hamamatsu R331-08
Operating temperature Ambient Ambient
Coincidence resolving time 22 ns 18 ns
Sum-coincident pulse amplification Logarithmic Linear
Pulse resolving time 5 μs to 33 μs
(variable with pulse height)
12μs
(fixed)
Spectral analog-to-digital converter (ADC) capacity 1000 channels 2048 channels
Nominal conversion gain (energy per channel) Variable
(with logarithmic energy)
≈1 keV
Detection threshold (nominal) ≤ 1 keV ≤ 1 keV
Live-time determination method (and uncertainty) Gated oscillator (scaled)
(± 0.1 %)
Gated oscillator (scaled)
(± 0.1 %)
Quench indicating parameter (QIP) Horrocks number (H) Transformed Spectral Index of the External Standard (tSIE) (proprietary)
External (-ray source for QIP determination 137Cs 133Ba
 (and location) (side) (bottom)