First Author [Ref], Institution, Publication year
|
Equipment
|
Image Guidance, Robotic Couch, Intrafraction Motion
|
Pre-RT Scans, Immobilization Devices
|
Patients
|
Technique, TPS
|
Notes, Results, or Conclusions
|
Mancosu [43]
Milan-Rozzano
2016
|
Varian Edge
120HD MLC
|
kV/CBCT
6-DOF couch
OSMS
|
CT
MRI
|
Phantom
|
-
|
Study of Edge linac with OSMS and CBCT. Tested ability of OSMS vs. CBCT ability to detect facial movements at isocenter, ability to recognize shifts, at different couch angles, and accuracy of OSMS when a camera is blocked. Submillimeter accuracy with rotational inaccuracy of 0.3 degrees.
|
Wen [44]
Henry Ford
2015
|
Varian Edge
120HD MLC
|
kV/CBCT
6-DOF couch
OSMS
|
-
|
Commissioning
|
FFF
VMAT
Cones
|
Report of commissioning of Edge radiosurgery system. Deviation between OSMS and CBCT was -0.4, 0.1, and 0 mm in vertical, longitudinal, and lateral dimensions. Beam data and mechanical parameters similar to TrueBeam, with advanced imaging package, 6-DOF couch, and intracranial SRS accessory package.
|
Seravalli [45]
MAASTRO
2015
|
Elekta Synergy
10 mm MLC
|
kV/CBCT
Pre- & post- CBCT
|
CT 1.2 mm
MRI 1.2 mm
Mask
(BlueBAG)
|
N = 52
Brain mets
|
SRS
Coplanar
VMAT
(Pinnacle)
|
Process of treatment. End-to-end test. GTV-PTV margin of 2.4 - 3.1 mm. Used Quantec constraints.
|
Li [29]
MSK
2015
|
Varian Trilogy
kV/CBCT
|
OSMS
|
Bite block
(PinPoint) vs
Open Mask (Freedom)
|
N = 25 Bite block
N = 8 Mask
|
FSRT
Coplanar beams
(iPlan)
|
Process of care diagram. Deliberate forced moves (15 volunteers) on ref Table 1. Study of volunteer comfort ref Table 2.
|
McTyre [40]
Wake Forest
2015
|
Gamma Knife Perfexion
|
No OBI
|
CT
MRI
Bite block (Extend)
|
N = 34
|
Fractionated GK
(GammaPlan)
|
Meningioma, schwannoma, metastases. GTV was treated without margin. 16-32 Gy to 50% isodose line over 4-5 fractions. Optic apparatus constrained to 4 Gy tangential to tumor. Daily repositioning errors < 1.2 mm.
|
Nanda [25]
Emory
2014
|
Novalis Tx
HD MLC
|
kV/CBCT
IR
6-DOF
|
CT 0.625 mm
MRI
|
N = 5
Pediatric
|
SRS/FSRT
Non-coplanar DCA
IMRT
12 beams
|
GTV-PTV 1 mm
4/5 patients required anesthesia
|
Pan [41]
UCSD
2012
|
TrueBeam
Trilogy
|
OSMS
|
CT 1.25 mm
MRI 1.25 mm
Open Mask
(CIVCO)
|
N = 44
Adults
|
SRS/FSRT
Multiple beams
Cones or VMAT
(Eclipse)
|
GTV-PTV 1 mm. Beam hold 1-2 mm and 1°. Treatment times – CBCT mean 11 min. Median shifts 1 mm, 2 mm, 1 mm vertical, longitudinal, lateral. Treatment time 15 min (shorter for TrueBeam). Compared local control to other series.
|
Schlesinger [39]
UVA
2012
|
Gamma Knife Perfexion
|
No OBI
|
CT
MRI
Bite block (Extend)
|
N = first 10
|
Fractionated GK
(GammaPlan)
|
Interfraction and intrafraction performance of Extend. Mean radial setup difference was 0.64 mm, SD 0.24 mm. Mean intrafractional positional difference was 0.47 mm. Cannot account for rotations.
|
Lu [35]
BIDMC
2012
|
Proton
|
Orthogonal kV
Three 2 mm gold fiducial spheres
|
CT
Frameless
|
N = 1 AVM
|
Proton
|
Description of novel technique with implanted fiducials to localize AVM identified on angiography and to transfer location information to CT for proton SRS planning.
|
Tryggestad [30]
JHU
2011
|
Elekta Synergy S
|
Pre- & post- CBCT
|
Mask - 4 types
Nonrandom study
Retrospective
|
N = 121
|
FSRT/external RT
|
Demonstrated masks (ref. Figure 1). Best was type-S head and shoulder mask with head and shoulder cushion with mouthpiece. Can achieve intrafraction motion of 1 mm or less, and interfraction variability of less than 3 mm.
|
Ramakrishna [31]
DFCI
2010
|
Novalis
|
Stereoscopic kV (ExacTrac)
IR
|
Frame
(Radionics)
Mask
(BrainLAB)
|
N = 102 SRS
N = 7 FSRT
|
SRS
|
End-to-end overall accuracy of Novalis Body ExacTrac is 0.7 mm ± 0.3 mm. Approximately 22% of mask-immobilized patients displayed intrafraction displacement of 1-2 mm.
|
Peng [49]
UF Gainesville
2010
|
Elekta Synergy
Varian Trilogy
|
|
CT 2 mm
Mask
IR
CBCT
|
N = 15 IR
N = 18 Mask
|
-
|
Comparison of IR tracking system setup with CBCT. Setup with IR resulted in setup errors of 1.2 mm determined by CBCT, versus mask and laser setup errors of 3.2 mm. FSRT should not rely on IR alone.
|
Mayo [34]
U Mass
2010
|
Varian Trilogy
5 mm MLC
|
kV/CBCT
|
CT 1.25 mm
MRI 1.25 mm
Mask
(Alpha Cradle)
|
N = 12
Adults
Brain mets
|
SRS
Non-coplanar VMAT
(Eclipse)
|
GTV-PTV 1-2 mm margin. Dosimetric details compared to CyberKnife, TomoTherapy, & IMRT. Reported on CI, gradient, & HI. Phantom end-to-end testing. Compared dose rate vs. survival in cell line (ref Figure 9).
|
Keshavarzi [50]
UCSD
2009
|
Varian Trilogy
|
IR
|
CT 1.25 mm
MRI 1.5 mm
Mask
(AccuForm)
|
N = 12
Pediatric
|
SRS/FSRT
MLC IMRT
Cones
(Eclipse)
|
GTV-PTV margin 1-3 mm
|