Table 5.

Summary of dose measurements using detector set 2.

Session SBDS 2 dose rate (µSv d−1) SBDS 2 dose >15 MeV (%) A07 dose rate (µSv d−1) A08 dose rate (µSv d−1)
A 14458+75 61 151±22 149±24
B 22819+45 100 150±22 153±25
C 12641+58 49 148±22 132±22
D 15415+41 100 187±26 166±26
E 16246+62 55 172±24 148±24
F n/a n/a 90±15 143±23
G n/a n/a 140±21 154±25
H n/a n/a 140±20 121±20
I n/a n/a 160±23 129±21
J n/a n/a 131±19 126±20
K n/a n/a 193±27 160±26
L n/a n/a 154±22 98±17
M n/a n/a 177±25 115±20

The data for sessions F–K were aggregated as described in the text. For sessions L and M, the data for SBDS 2 were excluded from the data analysis because of detector issues.