Skip to main content
. 2016 May 21;8(5):158. doi: 10.3390/toxins8050158

Table 5.

A summary of inter-run repeatability (RSD *) and accuracy of fortified kidney tissues using HPLC .

Fortification Levels, µg/g RSD *, % Average Nominal Concentration , µg/g Accuracy §, %
15.0 6.4 14.5 3.3
20.0 9.8 18.6 7.1
30.0 5.6 31.6 5.2
40.0 1.3 40.8 2.1
50.0 2.3 49.2 1.5

* Relative standard deviation. It is based on three analyses. High performance liquid chromatography. Nominal concentration is back-calculated from the calibration curve using the response of peak area from HPLC analysis. The average nominal concentration is calculated based on three analyses for each fortification level. § Accuracy is calculated as (Average nominal concentration-fortification level)/fortification level × 100%. It indicates the closeness of the concentration quantified by the calibration curve to the fortification level.