Table 2.
Water temperature measurement and control errors and standard uncertainties used in uncertainty analysis
Source of uncertainty | Range of possible temperature error values (mK) | Standard uncertainty (mK) |
---|---|---|
Electronic thermometer | ±1.0 | 1.0 |
Thermistor probe | ±1.5 | 1.5 |
Stability of bath | ±2.0 | 2.0 |
Bath setting | ±1.0 | 1.0 |
Est. immersion loss | ±2.0 | 2.0 |