Table 1.
Relationship of NIST and PTB units of luminous intensity (candela, cd) and luminous flux (lumen, lm) since 1985 CCPR intercomparison, relative to the world mean of 1985
Units | Participant | 1985 CCPR intercomp. | 1987 Bilateral intercomp. | 1988 BIPM report | 1990 New Int’n Temp. Scale | 1992 NIST New unit | 1993 Predicted |
---|---|---|---|---|---|---|---|
Candela | BIPM | 1.0100 | 1.000 | 1.000 | |||
PTB | 1.0032 | 1.003 | 1.003 | ||||
NIST | 0.9942 | 0.994 | 0.9977 | 1.0014 | 1.001 | ||
PTB–NIST | 0.0090 | 0.009 | 0.0055 | 0.0018 | 0.002 | ||
Lumen | BIPM | 0.9930 | 1.000 | 1.000 | |||
PTB | 1.0056 | 1.006 | 1.006 | ||||
NIST | 0.9899 | 0.990 | 0.990 | ||||
PTB–NIST | 0.0157 | 0.017 | 0.016 | 0.016 |