Table 2. Overall performance of RPA versus RT-PCR.

Sample type Sensitivity (%, 95% CI) Specificity (%, 95% CI) PPV (%) NPV (%) Positive LR Negative LR
Blood and swabs (n = 375) 264/271 (97, 95.5–99.3) 101/104 (97, 93.9–100) 99 94 33.8 0.03
Blood (n = 288) 213/219 (97, 95.1–99.4) 67/69 (97, 93.1–100) 99 92 33.6 0.03
Swabs (n = 87) 51/52 (98, 94.3–100) 34/35 (97, 91.6–100) 98 97 34.3 0.02