Skip to main content
. Author manuscript; available in PMC: 2016 Jun 1.
Published in final edited form as: Lasers Surg Med. 2015 Aug 19;47(8):658–668. doi: 10.1002/lsm.22383

TABLE 1.

Summary of Experimental Results

Trial Control setting Execution time (s) Frequency (target/s) Error (μm) Untreated rate Misaimed rate P-value between subjects
Automated 50 μm 46.55 0.69 45 ± 27 0.0% 0.0% 0.643
100 μm 26.25 1.22 50 ± 28 0.0% 0.0% 0.162
150 μm 21.75 1.47 57 ± 36 0.0% 0.0% 0.169
200 μm 18.72 1.74 60 ± 37 0.0% 0.0% 0.101
Manual 1.0 Hz 32.01 1.00 102 ± 67 5.2% 2.1% 0.003
1.5 Hz 21.35 1.50 124 ± 71 6.3% 4.7% 0.300
2.0 Hz 16.26 1.97 152 ± 92 11.5% 15.0% 0.411
2.5 Hz 13.20 2.42 174 ± 98 12.5% 20.1% 0.164