Table 5.2.
Relative standard uncertainties from all components during responsivity measurements at selected wavelengths
Component | Wavelength (nm) | AES | Relative standard uncertainty (%) | |||
---|---|---|---|---|---|---|
EPA | SERC | USDA | USDA | |||
Lamp | ||||||
Irradiance | 290 | 0.92 | 0.92 | 0.92 | ||
320 | 0.90 | 0.90 | 0.90 | 0.90 | 0.90 | |
350 | 0.71 | 0.71 | 0.71 | 0.71 | ||
Size | 0.09 | 0.09 | 0.03 | 0.01 | 0.01 | |
Goniometry | 0.46 | 0.46 | 0.27 | 0.11 | 0.11 | |
Current (random) | 290 | 0.06 | 0.06 | 0.06 | ||
320 | 0.05 | 0.05 | 0.05 | 0.05 | 0.05 | |
350 | 0.05 | 0.05 | 0.05 | 0.05 | ||
Current (systematic) | 290 | 0.11 | 0.11 | 0.11 | ||
320 | 0.10 | 0.10 | 0.10 | 0.10 | 0.10 | |
350 | 0.09 | 0.09 | 0.09 | 0.09 | ||
Alignment | 0.39 | 0.39 | 0.39 | 0.73 | 0.73 | |
Instrument | ||||||
Wavelength | 290 | 0.25 | 0.24 | 0.21 | ||
320 | 0.04 | 0.04 | 0.16 | 0.16 | 0.16 | |
350 | 0.05 | 0.03 | 0.12 | 0.12 | ||
Signal | 290 | 0.94 | 0.08 | 0.23 | ||
320 | 0.35 | 0.53 | 0.46 | 0.96 | 0.60 | |
350 | 0.18 | 0.07 | 0.85 | 0.41 | ||
Combined | ||||||
Random | 290 | 0.94 | 0.10 | 0.24 | ||
320 | 0.35 | 0.53 | 0.46 | 0.96 | 0.60 | |
350 | 0.19 | 0.09 | 0.85 | 0.41 | ||
Systematic | 290 | 1.14 | 1.13 | 1.06 | 1.19 | |
320 | 1.09 | 1.09 | 1.07 | 1.18 | 1.18 | |
350 | 0.94 | 0.94 | 1.04 | 1.04 |