. 2016 Mar 23;37(6):667–672. doi: 10.1017/ice.2016.35

FIGURE 1.

FIGURE 1

Mean ultraviolet C (UV-C) and standard deviation (SD) irradiance levels (W/cm 2 ) measured with 6 UV-C sensors placed at different distances and orientations relative to a mobile UV-C device in an intensive care unit room (left column at each position) and a room on a surgical ward (right columns). Sensor positions included in direct sight at distances of 1.3 m (4 ft) and 3.3 m (10 ft), at a 0° angle relative to the device at distances of 1.3 m and 3.3 m, and in shaded areas 1.3 m and 3.3 m from the device.