Table 5.

Summary of four large randomized trials on the use of pulse-oximetry for detection of CCHD.

COUNTRY GERMANY27 NORWAY28s SWEDEN26 UK29
Total infants 41,445 50,008 39,821 20,055
Study period 2006–2008 2005–2006 2004–2007 2008–2009
Age at screening 24–72 hrs 1–21 hrs 1–406 hrs Before discharge
Pulse oximetry site Postductal Postductal Pre and postductal Pre and postductal
Oxygen saturation cutoff ≥96% ≥95% ≥95% ≥95%
Pre – post ductal oxygen saturation difference Not applicable Not applicable >3% >2%
Number of retest 1, after 1 hr 1, after 2–3 hr 2, 1 hr apart 1, after 1–2 hr
Screening staff Routine care providers Routine care providers Routine care providers Routine care providers
Equipment variable Standardized Standardized Standardized
Sensitivity 77.8 77.1 62.07 75
Specificity 99.9 99.4 99.82 99.12
Positive predictive value 25.9 8.3 20.69 9.23
Negative predictive value 99.99 99.98 99.97 99.97
False positive rate 0.10 0.6 0.17 0.8