Skip to main content
. 2015 Nov 7;172:37–45. doi: 10.1007/s12011-015-0557-3

Table 2.

Results of the measurements (n = 10) of certified reference materials: SLRS-5 and TM 27.3 given with certified values

Analyte River water SLRS-5 Fortified lake water TM 27.3
Measured value (μg L−1) Certified value (μg L−1) Accuracy (%) Measured value (μg L−1) Certified value (μg L−1) Accuracy (%)
Al 46.84 ± 0.65 49.5 ± 5.0 95 42.8 ± 1.7 44.4 ± 5.1 96
As 0.448 ± 0.039 0.413 ± 0.039 109 2.258 ± 0.030 2.15 ± 0.30 105
Ba 14.6 ± 1.3 14.0 ± 0.5 104 15.96 ± 0.21 14.9 ± 1.1 107
Cd 0.0095 ± 0.0021 0.0060 ± 0.0014 159 1.110 ± 0.014 1.05 ± 0.12 106
Co 0.0607 ± 0.0028 0.05a 121 1.888 ± 0.020 2.05 ± 0.18 92
Cr 0.289 ± 0.044 0.208 ± 0.023 139 1.953 ± 0.010 1.73 ± 0.33 113
Cu 17.07 ± 0.54 17.4 ± 1.3 98 6.01 ± 0.12 6.16 ± 0.61 98
Mg 2752 ± 46 2540 ± 160 108
Mn 4.12 ± 0.32 4.33 ± 0.18 95 2.232 ± 0.016 2.27 ± 0.35 98
Ni 0.589 ± 0.016 0.476 ± 0.064 124 2.396 ± 0.070 2.44 ± 0.53 98
Sr 52.40 ± 0.65 53.6 ± 1.3 98 98.4 ± 1.4 105 ± 6 94
U 0.0908 ± 0.0090 0.093 ± 0.006 98 2.094 ± 0.075 2.03 ± 0.19 103
V 0.360 ± 0.049 0.317 ± 0.033 114 2.170 ± 0.042 2.18 ± 0.25 100

ainformation value