Table 1. Analytical sensitivity of the MG 23S assay as determined by detection of replicate of different concentration of each template.
Copy number Detected/replicate tested | Copy number (detected/ replicates tested) | ||||||
---|---|---|---|---|---|---|---|
Template | 105 | 104 | 103 | 102 | 10 | 1 | |
A2058G | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 2/3 | 12 (20/20) |
A2058C | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 10 (20/20) |
A2058T | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 1/3 | 10 (19/20) |
A2059G | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 1/3 | 12 (20/20) |
A2059C | 3/3 | 3/3 | 3/3 | 3/3 | 2/3 | 0/3 | 15 (20/20) |
MgPa | 3/3 | 3/3 | 3/3 | 3/3 | 3/3 | 2/3 | 10 (20/20) |