Table 2.

Deviation from “trueness”: Reference mean values and devices’ mean differences from reference [mg/dL (SD)] for WB and P, for the whole range and per glycemic interval.

Device Blood Source Mean difference with reference (SD)
Whole Range (N = 100) Hypoglycemia (N = 15) Normoglycemia (N = 38) Hyperglycemia (N = 47)
Reference value Mean (SD) 175.30 (115.74) 51.38 (14.98) 106.15 (22.30) 265.04 (106.39)
Aviva WB 19.32 (28.19) 1.38 (12.61)d 8.87 (13.05) 32.67 (33.67)d
P −10.90 (15.20) −6.54 (3.48) −5.69 (6.40) −16.31 (19.82)
FreeStyle WB 74.15 (55.24) 18.69 (13.60) 40.38 (21.09) 116.60 (48.83)
P 0.89 (20.48)b −2.00 (4.95)d 1.49 (9.85)d 1.19 (28.21)d
Glucocard WB 48.18 (30.84) 25.54 (10.62) 40.64 (11.60) 60.44 (39.02)
P 28.10 (20.21) 15.46 (5.97) 24.97 (6.62) 34.06 (26.97)
Hemocue WB 47.83 (49.15)a 2.75 (20.93)d 22.08 (15.92) 80.02 (51.00)
P −10.79 (32.66) −19.15 (9.61) −18.49 (7.85) −2.27 (45.07)d
Ultra WB 38.7 (30.99) 16.92 (13.16) 27.15 (13.53) 54.23 (36.56)
P −64.71 (43.32) −6.62 (11.41)d −40.87 (17.16) −99.81 (31.91)
Verio WB 29.96 (32.45)a 6.77 (11.19)d 14.44 (10.09) 49.25 (37.28)
P −13.19 (14.42)a −5.42 (5.47) −8.38 (7.16) −19.17 (17.81)
Vita WB 38.82 (30.95) 16.46 (12.69) 25.15 (11.56) 54.58 (36.25)
P −59.61 (40.45) −3.61 (9.61)d −35.80 (13.71) −94.13 (26.76)
Optium WB 38.57 (34.76) 10.85 (14.96) 19.03 (15.32) 61.96 (34.78)
P −48.56 (31.95) −13.46 (9.67) −39.00 (11.99) −65.83 (35.90)
StatStrip WB 33.28 (30.28) 12.62 (11.09) 20.44 (15.21) 49.26 (34.57)
P −3.92 (26.87)c 4.85 (4.78) 5.62 (7.32) −14.04 (35.64)

WB, whole blood; P, plasma; PBGMs, portable blood glucose meters.

a One hundred determinations were obtained for all devices, except Verio (99 WB and 98 P samples) and Hemocue (99 WB samples). Normoglycemia is defined between 73 and 143 mg/dL.

b, cAll PBGMs showed significant differences with the reference in both whole blood and plasma, for the whole glucose range, with the exception of bFreestyle (P = .665) and cStatStrip (P = .148) in plasma samples, and Vita and Ultra for the same samples in the hypoglycemic range (P = .200 and P = .059, respectively) and Hemocue (P = .729), in hyperglycemia.

d P > .05 (nonsignificant) compared with reference by glycemic ranges.