Table 3.

CTDIvol(32 cm) for four different tube potentials and a tube-current-time-product of 100 mAs. CTDIvol values for higher tube-current-time-products scale linearly. The small difference between the two systems is due to the different detector collimations used (EID, 32 × 1.2 mm; PCD, 32 × 0.5 mm). The EID detector’s total detector coverage is greater, hence the CTDIvol is about 10% less, as wider collimation settings are more dose efficient.

Tube Potential (kV) EID CTDIvol (mGy) PCD CTDIvol (mGy)
80 2.0 2.2
100 4.2 4.6
120 6.9 7.6
140 10.1 11.1