Published in final edited form as: Invest Radiol. 2016 Jul;51(7):421–429. doi: 10.1097/RLI.0000000000000251

Table 4.

Acquisition and reconstruction parameters used to scan the whole body cadaver. All parameters were identical for the head and body scans, unless otherwise noted.

Parameter Subsystem A (EID) Subsystem B (PCD)
Detector collimation 32 × 1.2 mm (thorax)
32 × 1.2 mm (abd/pel)
64 × 0.6 mm (legs)
32 × 0.5 mm
Tube potential (kV) 140 140
Tube current (mA) – head not performed 400
Tube current (mA) – body 180 (thorax)
240 (abd/pel)
250 (legs)
180 (thorax)
240 (abd/pel)
250 (legs)
CTDIvol (mGy) – head, 16 cm not performed 101.4
CTDIvol (mGy) – body, 32 cm 17.7 (thorax)
23.6 (abd/pel)
24.6 (legs)
20.0 (thorax)
26.6 (abd/pel)
27.7 (legs)
Rotation time (s) 1.0 1.0
Acquisition FOV (mm) 500 275
Scan mode Spiral Spiral
Spiral pitch 1.2 (thorax)
0.6 (abd/pel)
0.8 (legs)
0.5
Image thickness (mm) 5.0 5.0
Image interval (mm) 5.0 5.0
Reconstruction kernel D30 D30
Recon. FOV (mm) – head 200 200
Recon. FOV (mm) – body 275 275