Table 4.
Parameter | Subsystem A (EID) | Subsystem B (PCD) |
---|---|---|
Detector collimation | 32 × 1.2 mm (thorax) 32 × 1.2 mm (abd/pel) 64 × 0.6 mm (legs) |
32 × 0.5 mm |
Tube potential (kV) | 140 | 140 |
Tube current (mA) – head | not performed | 400 |
Tube current (mA) – body | 180 (thorax) 240 (abd/pel) 250 (legs) |
180 (thorax) 240 (abd/pel) 250 (legs) |
CTDIvol (mGy) – head, 16 cm | not performed | 101.4 |
CTDIvol (mGy) – body, 32 cm | 17.7 (thorax) 23.6 (abd/pel) 24.6 (legs) |
20.0 (thorax) 26.6 (abd/pel) 27.7 (legs) |
Rotation time (s) | 1.0 | 1.0 |
Acquisition FOV (mm) | 500 | 275 |
Scan mode | Spiral | Spiral |
Spiral pitch | 1.2 (thorax) 0.6 (abd/pel) 0.8 (legs) |
0.5 |
Image thickness (mm) | 5.0 | 5.0 |
Image interval (mm) | 5.0 | 5.0 |
Reconstruction kernel | D30 | D30 |
Recon. FOV (mm) – head | 200 | 200 |
Recon. FOV (mm) – body | 275 | 275 |