Table 5.

Acquisition and reconstruction parameters used to scan the isolated cadaveric head.

Parameter Subsystem A (EID) Subsystem B (PCD)
Detector collimation 64 × 0.6 mm 32 × 0.5 mm
Tube potential (kV) 140
Tube current (mA) 550 550
CTDIvol (mGy) – Head, 16 cm 124.5 139.4
Rotation time (s) – head 1.0
Acquisition FOV (mm) 500 275
Scan mode Spiral
Spiral pitch 0.6
Image thickness (mm) 5.0
Image interval (mm) 5.0
Reconstruction kernel D40
Reconstruction FOV (mm) 200