Skip to main content
. Author manuscript; available in PMC: 2017 Jun 1.
Published in final edited form as: Semin Fetal Neonatal Med. 2016 Feb 18;21(3):162–173. doi: 10.1016/j.siny.2016.02.001

Table 1.

Potential ventilator driver, nasal interface options and support features for initiation of neonatal high frequency nasal ventilation.

Variable Comments
Ventilator A variety of “drivers” are possible – see Table 7
Not all have been used clinically or studied in the lab
Frequency Dependent on device & active v passive expiratory phase
Optimum f unclear
Recommendation:
  Start oscillators at 6 – 8 Hz; others at 4 – 6 Hz
Inspiratory time May be expressed as I:E, I-time or “On-time”
VT ≫ at 50% I:E compared to 33%
Recommendation:
  Set I:E at 50%; for Jet use longer “on-time” than 20 msec
Amplitude/ΔP Device dependent
Increased ΔP → larger VT ; appears to plateau ~ 70%
Recommendations:
  Start ~ 50% max ΔP for device
  Adjust as possible to achieve visible chest wall vibration
Nasal interface Single naso-pharyngeal tube
Standard binasal CPAP prongs or nasal CPAP mask
? other nasal cannula interfaces
Recommendations:
  Maximize internal diameter → Larger VT
  Minimize dead space
Conventional breaths Not all devices can provide this additional support
Studied in animal studies, but not described in neonatal reports