Table 5.
Uncertainty budget for luminous intensity measurements
| Source of uncertainty | Relative standard uncertainty (%) |
||
|---|---|---|---|
| Type A | Type B | ||
| Illuminance Responsivity | |||
| Scale uncertainty from Table 2 | 0.19 | ||
| Measurement noise | 0.02 | ||
| Filter absorption | 0.006 | ||
| Lamp to Photometer Distance | |||
| Size and construction of lamp | 0.06 | ||
| Physical distance measurement | 0.01 | ||
| Geometrical | |||
| Photometer transverse placement | a | ||
| Photometer orthogonality | 0.002 | ||
| Lamp Operation | |||
| Current regulation | 0.03 | ||
| Aging (per hour) | 0.1 | ||
|
| |||
| Combined standard uncertainty | 0.23 | ||
| Expanded uncertainty (k = 2) | 0.46 | ||
Too small to list.